Pyrene
fluorescence assay of actin assembly
Dyche Mullins, 9/13/99
Stock solutions:
Buffer A (a.k.a. G buffer):
2 mM Tris HCl, pH 8.0
0.2 mM ATP
0.1 mM CaCl2
0.5 mM DTT
KMEI (Polymerization
buffer):
50 mM KCl
1 mM MgCl2
1 mM EGTA
10 mM Imidazole HCl, pH 7.0
ME (Mg-exchange buffer)
50 µM MgCl2
0.2 mM EGTA
Useful facts:
Extinction
coeff of actin at 290 nm: 38.5µM/O.D.
Correction
for pyrene actin - subtract 0.127´A344
from A290
Extinction
coeff of pyrene at 344 nm: 45.0µM/O.D.
PREFERRED
WAVELENGTHS:
Excitation:
365 nm
Emission: 407 nm
ACCEPTABLE WAVELENGTHS:
Excitation: 365 nm
Emission: 385 nm
Before starting:
1. Be sure you have good reagents.
a.
Actin must be gel-filtered and you should only use fractions from the back
side of the gel filtration peak.
b.
You can use 100% pyrene labeled actin but be careful. Not all actin-binding proteins bind as tightly to pyrene actin
(e.g. profilin!). The perferred range
of pyrene doping is 1-10%.
c.
Don't use frozen actin for nucleation assays, it assembles with measurably
slower kinetics.
d.
Don't use old actin. Monomeric actin
doesn't store very well. If you're
particularly conservative don't use actin more than two weeks old. If you're not so conservative you can use it
for 4-6 weeks but be aware that you are pushing your luck. Pyrene actin keeps better than dark actin so
you can use quite old (several months old) pyrene actin with fresh dark actin.
e.
Check your polymerization curves versus a standard curve. If you have an
untried batch of actin, do a spontaneous polymerization assay at a known
concentration and compare it to the curves attached to this sheet. If there is
a big difference then either i) your concentration is off or ii) your actin is
bad. Do not proceed until you have figured this out. Wasting time is bad!
2.
Turn on the fluorimeter and fire up the computer. Make sure everything is turned
off at the beginning. The correct
sequence of events for the ISS K2 is:
a.
Turn on the arc lamp - This is the important part. The lamp is a 300 W xenon
arc lamp and it can kill nearby electronic equipment and unsupervised
children. Turn the current control knob
all the way to the left (minimum current) and flip the switch. Wait until you
hear the audible pop of the arc starting and/or until you see the panel meter
read 10 Amps.
b.
Turn on the fluorimeter. The switch is in the right rear of the
instrument. Make sure the PMT you want
is selected and set to the correct voltage.
For most applications, use the PMT on the right hand side. Set the toggle on the PMT to
"manual" and dial up a voltage between 900-970V.
c.
Turn on the computer. Make sure to switch on both the monitor and CPU. Wait for the computer to boot up and then,
at the prompt, type "iss".
For polymerization assays use the "Take a Spectrum" option
(#2). Then choose "Intensity" (option #2 again) and "Slow
Kinetics" (also option #2).
d.
Finally bring the lamp up to operating voltage. Slowly turn the current knob
until the panel meter reads 15 Amps.
e.
The shut-down procedure is the exact reverse of the above sequence.
3.
Optimize the fluorimeter. You want to maximize signal and minimize light
scattering and photobleaching.
a.
Photobleaching. To minimize photobleaching, you must minimize the excitation
light. Settings that have worked for me
are: 0.5 mm excitation slits, excitation polarizer in line (this also seems to
help with light scattering) and excitation iris closed 80-90%. Remember, you
cannot simply correct for photobleaching mathematically. There are free
radicals involved. Very dangerous.
b.
Light scattering. Large structures
(aggregates, vesicles etc.) scatter incident light. This causes a problem because the monochrometers on the spec
aren't perfect and will pass a small amount of light at the excitation
frequency. You can cut down on light
scattering by spinning and degassing your sample before use, by putting the
excitation polarizer in line (the monochrometers are more selective when
dealing with vertically polarized light), by using the smallest monochrometer
slits possible, and by using a low-pass filter on the emission channel (the 399
nm cutoff works well for pyrene fluorescence).
c.
Signal. If you don't have enough
signal, or if your signal to noise ratio (SNR) is crappy, you can do several
things: i) increase the ratio of pyrene:dark actin, ii) increase the emission
slit width, iii) increase the PMT voltage (don't go up to 1000 volts), iv)
increase the excitation light by opening the iris or increasing the slit widths
or v) as a desperate measure, you can use the left emission channel (this
channel has no monochrometer so the light throughput is much higher. You just
need a good low pass cutoff filter).
d.
Check your settings! Polymerize 100 uL
of your stock solution of pyrene actin then dilute it down to working concentration
and load it into the fluorimeter.
Collect data for several hundred seconds. If the signal is trending downward, you are photobleaching and
you should reduce your excitation light.
If your signal is too strong you should also decrease the size of the
excitation and/or emission slits. If the signal is too weak, see step (c)
above.
Tips
for collecting good data:
1.
Polymerize by adding 0.1 volumes of 10X KMEI.
This buffer gives the optimum polymerization rate (See Drenkenham and
Pollard, 1984 J. Biol. Chem. for
details) and also approximates the intracellular ionic composition. Bring all of your stock solutions together
and triturate several times for good mixing. Schlieren lines are your enemy!
2.
Pre-exchange the calcium for magnesium.
Monomeric actin is stored in calcium and if you polymerize it by simply
adding 10X KMEI, the Ca-Mg exchange shows up in the kinetics. To pre-exchange the Mg, incubate the
pyrene-doped actin stock solution with 0.1 vol. of 10X ME (see above) for about
2 minutes before adding the KMEI. Do
not pre-exchange all of your actin at once. It will polymerize before you can do the assays!
3.
Keep your buffers at room temperature.
If your sample starts out cold at time zero, and warms to room
temperature during data collection, the kinetics will be way off. If you are adding a large volume of buffer
to small volumes of protein (say 10:1), you can keep the protein on ice and the
buffer at R.T.
4.
Keep track of the dead time. The ISS
does an admirable but lamentably slow dark count correction for each data
set. To collect analyzable data use a
stop watch. Start it as soon as you
begin the polymerization reaction and stop it as soon as the first data point
shows up on the screen. Record the dead
time for each data set and add it back to the X-axis data before analysis. In this manner you may more quickly approach
the way of the kinetic warrior.
5.
Watch for air bubbles. When loading,
tilt the 60 µL quartz cuvette to one side and put your pipette tip in one
corner. Fill slowly and watch for
bubbles. If you get a bubble, withdraw
the liquid and try again. As long as you're measuring dead time with a
stopwatch it's okay. Take a deep breath. Try again.
Bubbles kill data dead!
6.
Use a fresh box of pipette tips and keep the lid closed when not in use. Dust is bad.
7.
Wash the cuvette thoroughly between each run. Little bits of leftover actin
will nucleate filament formation in your next run. Use the cuvette washer and don't spare the water. Rinse water, ethanol, water - both inside and out. Wipe down the outside to remove
fingerprints. Dry the cuvette on the washer before reuse.
8.
Collect complete data sets from time
zero until the pyrene fluorescence plateaus.
You cannot compare or analyze partial data sets.
9.
If you have determined that the conditions you are using don't alter the
critical concentration, you can normalize your data. First, subtract the baseline values (fluorescence at time zero)
from each data set. Then divide all the
data in each dataset by the fluorescence at the plateau. This will correct for small pipetting errors
and variations in the cuvette position.
10.
Make your mind right. Remember what
Henry V says: "All else be so, if our minds
be right."